The Business Centre Medical (CAM)<br/>In a space of 2000 m2, with more than 7,000 articles are presented, stored, inspected, refurbished or reconditioned.<br/>Our offer covers all the needs of the profession, at all levels : equipment and medical devices, of course, but also technical equipment and furniture, for offices, clinics or hospitals.<br/>Why pay more ?<br/>The Business Centre Medical (CAM), is a company specialized in the sale of all types of equipment, medical, of which the common denominator is that they are for sale at discount prices. To meet this economic criteria, CAM responds to the needs for the two divisions :<br/>Department of new equipment<br/>Department equipment refurbished
Description
CALIBRATION
Vaporizers are calibrated at 21°C using oxygen carrier gas at a flow rate of 5 liters/minute and are temperature, flow, and pressure compensated within the specified operating range.
GAS FLOW RESISTANCE WITH VAPORIZER OFF
Resistance Vaporizer Setting Flow Rate Temperature Temperature
10 -15 cm wg ON + 0% 5 liters/minute, oxygen 21 ± 2°C
LIQUID CAPACITY
The amount of liquid required to fully charge a vaporizer with dry wicks = 300 milliliters (nominal).
The amount retained by the wick system = 75 milliliters (nominal).
5% Vaporizer
The amount of liquid to fill the vaporizer from minimum to maximum = 170 milliliters (nominal).
8% Vaporizer (Sevotec)
The amount of liquid to fill the vaporizer from minimum to maximum = 137 milliliters (nominal).
WEIGHT AND DIMENSIONS
ENVIRONMENTAL CONDITIONS
Operating Temperature Range = 18°C to 35°C
Storage Temperature Range = –40°C to +65°C
PERFORMANCE
Accuracy to 5 liters/minute of O2 at 21°C ± 2°C:
Vaporizer 5% ± 0.25% of delivered agent or
± 15% of dial setting (whichever is greater).
Vaporizer 8% ± 0.4% of delivered agent or
± 15% of dial setting (whichever is greater)