
Product Details
| Model | DK-TMS |
| Brand | DUKON |
| Category | Transcranial magnetic stimulators |
| Posted | 05.02.2026 |
| Views | 822 |
New DUKON DK-TMS
Safe deal
Bimedis secures your payment until you confirm the equipment is received.
Description
Product Peformance
(1) Five modes of output:
a) Mode 1 (Direct Current Therapy Wave) has a maximum output current of 2mA with an error
of±20 %, and selectable steps of 0.02mA and 0.1mA.
b) Mode 2 (Triangular Therapy Wave) has a maximum output current of 2mA with an error of
±20%, and selectable steps of 0.02mA and 0.1mA.
c) Mode 3 (Sinusoidal Therapy Wave) has a maximum output current of 2mA with an error of
±20%, and selectable steps of 0.02mA and 0.1mA.
d)Mode 4 (Pulse Therapy Wave) has a maximum output current of 15mA with an error of ±20%,
and selectable steps of 0.5mA and 0.1mA.
e)Mode 5 (Sham Stimulation) has no output.
(2) Maximum Load Capacity: When the output maximum load is 20kΩ, the current is adjusted to the
maximum, and the output voltage is 40V with an error of ±20%.
(3) Constant Current Output Characteristics: The output current varies within an error of ±20% under
loads from 0 to 20kΩ.
(4) Incremental Adjustment Characteristics:
Modes 1/2/3: The current is modulated in two increments of 0.02mA and 0.1mA.
Mode 4: The current is modulated in two increments of 0.1mA and 0.5mA.
(5) Therapy Waveform Frequency:
Modes 2/3 have a therapy waveform frequency of 0.5Hz with an error of ±10%.
Mode 4 (Pulse Waveform) has a therapy frequency of 0.25Hz with an error of ±10%.
(6) Therapy Time Selection: Adjustable within the range of 5 to 60 minutes, with the smallest adjustable
time interval being 5 minutes.
(7)Electrode Detachment Detection: Within 10 seconds of electrode disconnection, there will be a voice
prompt indicating electrode detachment and the end of treatment.
(8)Automatic Power-off Function: The device will automatically power off after 5 minutes of inactivity.
(9)Lead Disconnection Output Protection: If the lead disconnects, the output voltage will automatically be
cut off.
(10)Low Battery Indicator: The device will provide a low battery warning when the battery level is too low.
(11)Output Control: When initiating treatment, it should always start from 0mA, otherwise, the electrical
stimulator should not have output.
(12)Open-Circuit Voltage Output: When the stimulator output is open-circuited, the measured open-circuit
voltage at the output terminals should not exceed 50V.
Didn’t find what you were looking for? Place your own ad on the Bimedis platform
Place your advert on Bimedis and reach buyers worldwide.