ModelDK-TMS
BrandDUKON
CategoryTranscranial magnetic stimulators
Posted05.02.2026
Views822

New DUKON DK-TMS


Condition:: New
Warranty (month):: 12
Year:: 2026
Availability:: In stock
Shipping from:: China
Sales area:: Afghanistan, Algeria, American Samoa

41 000

Safe deal

Bimedis secures your payment until you confirm the equipment is received.

Description

Product Peformance

(1) Five modes of output:

a) Mode 1 (Direct Current Therapy Wave) has a maximum output current of 2mA with an error

of±20 %, and selectable steps of 0.02mA and 0.1mA.

b) Mode 2 (Triangular Therapy Wave) has a maximum output current of 2mA with an error of

±20%, and selectable steps of 0.02mA and 0.1mA.

c) Mode 3 (Sinusoidal Therapy Wave) has a maximum output current of 2mA with an error of

±20%, and selectable steps of 0.02mA and 0.1mA.

d)Mode 4 (Pulse Therapy Wave) has a maximum output current of 15mA with an error of ±20%,

and selectable steps of 0.5mA and 0.1mA.

e)Mode 5 (Sham Stimulation) has no output.

(2) Maximum Load Capacity: When the output maximum load is 20kΩ, the current is adjusted to the

maximum, and the output voltage is 40V with an error of ±20%.

(3) Constant Current Output Characteristics: The output current varies within an error of ±20% under

loads from 0 to 20kΩ.

(4) Incremental Adjustment Characteristics:

Modes 1/2/3: The current is modulated in two increments of 0.02mA and 0.1mA.

Mode 4: The current is modulated in two increments of 0.1mA and 0.5mA.

(5) Therapy Waveform Frequency:

Modes 2/3 have a therapy waveform frequency of 0.5Hz with an error of ±10%.

Mode 4 (Pulse Waveform) has a therapy frequency of 0.25Hz with an error of ±10%.

(6) Therapy Time Selection: Adjustable within the range of 5 to 60 minutes, with the smallest adjustable

time interval being 5 minutes.

(7)Electrode Detachment Detection: Within 10 seconds of electrode disconnection, there will be a voice

prompt indicating electrode detachment and the end of treatment.

(8)Automatic Power-off Function: The device will automatically power off after 5 minutes of inactivity.

(9)Lead Disconnection Output Protection: If the lead disconnects, the output voltage will automatically be

cut off.

(10)Low Battery Indicator: The device will provide a low battery warning when the battery level is too low.

(11)Output Control: When initiating treatment, it should always start from 0mA, otherwise, the electrical

stimulator should not have output.

(12)Open-Circuit Voltage Output: When the stimulator output is open-circuited, the measured open-circuit

voltage at the output terminals should not exceed 50V.