- Type of analyzer
- closed
- Throughput, samples per hour
- 36
- Analizator type
- Automatic
- Sample Types
- Plasma, Serum
- Cycle time (reaction), s
- 96
- Weight (kg)
- 40
Used BIOMERIEUX Mini VIDAS
Condition:: Used
Availability:: In stock
Shipping from::
Poland
Safe deal
Bimedis secures your payment until you confirm the equipment is received.
Description
Power supply 100-240VAC. Software: R5.3.0, Firmware 1.13, Scanner version: 2.52, Section A and B: 3.10, composition of III generations. Revealing - replaceable with a new one (MK). Comprehensive external and internal maintenance. Mechanics inspection, oiling and lubrication. Rubber maintenance. Optics maintenance. Header removal - was not. Handheld scanner - functional. Home position: 5 x 0 degrees. Control: pump (see below). OPT and QCV control - see catalog for repeatability. During QCV, one strip was not read to position 2, then 1, was read to position 6. After printing the QCV results of section B, the QCV test of section A was repeated, in the initial test release the analyzer was verified. critical error: Note: A power failure occurred while the test was being performed. Results may have been lost.. The QCV test in section A was repeated. The analyzer seems to be effective. /// 2023.0112 Upgrade (JS) to the latest version: R5.6.1 and introduced tests: LEAK, CDAB, RBG, TES2.