Type of analyzer
closed
Number of parameters
18
Sample size, microlitre
40
Throughput, samples per hour
80
Analizator type
Automatic
Sample Types
Whole blood
Cycle time (reaction), s
45

Used SYSMEX K-4500


Condition:: Used
Availability:: In stock
Shipping from:: Poland

Safe deal

Bimedis secures your payment until you confirm the equipment is received.

Description

Analyzer without feeder. Power supply 220-230VAC. The measuring block (there was a leak) along with the aluminum housing and the printer with the electronic module were exchanged with another analyzer (SN: AB429) (the printer failed to print the data several times). He had problems with repeatability (blood problem?, insulated measuring electrode?, dirty orifice?). The final repeatability measurement was in line with expectations. The analyzer is operational.