Product Details
| Model | BC-5800 |
| Brand | MINDRAY |
| Category | Hematology Analyzer |
| Posted | 26.01.2026 |
| Views | 322 |
Product Specifications
- Overall dimensions, cm ⓘ
- 66,4 х 61,3 х 58,5
- Type of analyzer
- open
- Number of parameters
- 27
- Sample size, microlitre
- 40-140
- Throughput, samples per hour
- 90
- Analizator type
- Automatic
- Sample Types
- Venous blood, Capillary blood
- Cycle time (reaction), s
- 40
- Weight (kg)
- 76
Used MINDRAY BC-5800
Condition:: Used
Year:: 2015
Availability:: In stock
Shipping from::
Poland
Safe deal
Bimedis secures your payment until you confirm the equipment is received.
Description
BC-5800, Mindray, SN: RW-57003527
Year of manufacture 07.2015. 100-240VAC power supply for analyser. Compressor power supply 220/230VAC.Maintenance, removal of labels.All tests performed without reagents.Analyser mechanical/electrical/electronic side starts correctly.
Test procedure:
1. The presence of reagents was simulated.
2. expiry date of reagents was changed.
3. analyser software version - Boot software:1.3.5
4. Temp&Pres - temperatures and pressures within correct ranges.
Vol&Cur - voltages and currents in correct ranges.Laser diode current: 29.7 (range 20-50).
6. Self test: tests (Mechanical, Circuit, Valve) positive except:RBC Aperture Vol, Hgb Blank Voltage. Dobber - no reagent. Errors are probably due to lack of reagents.
Setup: date correction, Print - delete user data, Advanced - change language from Polish to English.
Kit composition:
1. analyser
2. compressor, year of manufacture 07.2015, SN: RWA57013627, 220/230VAC.
3. hand-held barcode scanner (Newland, NLS-HR100, SN: H55E13566NOM)
4. complete hosing and cabling
6 Sampling wands 6 pcs.
7. feeder, SN: RW-57003527.
Year of manufacture 07.2015. 100-240VAC power supply for analyser. Compressor power supply 220/230VAC.Maintenance, removal of labels.All tests performed without reagents.Analyser mechanical/electrical/electronic side starts correctly.
Test procedure:
1. The presence of reagents was simulated.
2. expiry date of reagents was changed.
3. analyser software version - Boot software:1.3.5
4. Temp&Pres - temperatures and pressures within correct ranges.
Vol&Cur - voltages and currents in correct ranges.Laser diode current: 29.7 (range 20-50).
6. Self test: tests (Mechanical, Circuit, Valve) positive except:RBC Aperture Vol, Hgb Blank Voltage. Dobber - no reagent. Errors are probably due to lack of reagents.
Setup: date correction, Print - delete user data, Advanced - change language from Polish to English.
Kit composition:
1. analyser
2. compressor, year of manufacture 07.2015, SN: RWA57013627, 220/230VAC.
3. hand-held barcode scanner (Newland, NLS-HR100, SN: H55E13566NOM)
4. complete hosing and cabling
6 Sampling wands 6 pcs.
7. feeder, SN: RW-57003527.