Bimedis

Product Details

ModelBC-5800
BrandMINDRAY
CategoryHematology Analyzer
Posted26.01.2026
Views304

Product Specifications

Overall dimensions, cm
66,4 х 61,3 х 58,5
Type of analyzer
open
Number of parameters
27
Sample size, microlitre
40-140
Throughput, samples per hour
90
Analizator type
Automatic
Sample Types
Venous blood, Capillary blood
Cycle time (reaction), s
40
Weight (kg)
76

Used MINDRAY BC-5800


Condition:: Used
Year:: 2015
Availability:: In stock
Shipping from:: Poland

Safe deal

Bimedis secures your payment until you confirm the equipment is received.

Description

BC-5800_SN_RW-56003521
Year of manufacture 06.2015. Power supply for analyser 100-240VAC, 300VA. Maintenance, removal of labels. Tests performed: without reagents, without compressor, without reagent connections. Analyser mechanical/electrical/electronic side starts correctly - except for the deficiencies described below. Password: ADMIN/ADMIN. Selected mechanical tests - positive. Temperatures - correct. Power supply voltages - correct. Laser diode current: 32.13 (range 20-50). Language change from Polish to English. Boot software version of analyser: 1.3.5
LACKS:
1. Feeder missing.
2. ‘Pneumatic Unit’ compressor missing.
3. Multiple reagent connections, hoses, effluent sensor with oviposition, power cable missing.
4. CR1220 battery damaged, replacement necessary, analyser loses date and time.
5. 2 screws of the rear sheet metal broken.
6. Pressure connection at rear of analyser missing.
Broken plastic of the front casing. 8.
8. Hand-held barcode reader.

Didn’t find what you were looking for? Place your own ad on the Bimedis platform

Place your advert on Bimedis and reach buyers worldwide.

SELL