Product Specifications

Measured points
1500
Measurement range, D
1-100
Wavelength, nm
555
Measurement technology
Placido rings (22)
Accuracy, D
± 0.05

For Parts ZEISS i.Profiler Plus


Condition:: For Parts
Availability:: In stock
Shipping from:: USA
Sales area:: Afghanistan, Albania, Algeria

5 000

Safe deal

Bimedis secures your payment until you confirm the equipment is received.

Description

This device is listed for part because it shows a chin rest error.
This device is listed for part because it shows a chin rest error. It will be sold as is without warranty or return.

ZEISS iProfiler Plus Autorefractometer, ATLAS Corneal Topographer, Aberrometer, and Keratometer
The iProfiler plus is the 4-in-1 compact system with ocular wavefront aberrometer, autorefractometer, ATLAS corneal topographer and keratometer. The fully automated measurement procedure, with easy-to-use touch screen control, enables all measurements of both eyes in approximately 60 seconds. The eye’s refractive power distribution is analysed and represented across the entire pupil aperture.

High-resolution Hartmann-Shack wavefront sensor. The wavefront is sampled at 1,500 points across 7 millimetre pupil aperture
Adjustable chin rest for comfortable head positioning
Fully automated measurement procedure
i.Profilerplus features a high- resolution wavefront measurement and corneal topography which provide you with all data needed in order to evaluate the refractive status of your patient´s eyes.
The analysis mode enables you to visualize the impact of different aberrations on your patient´s vision, including up to 7th Zernike aberrations. Additionally, the benefits of i.Scription® technology can be simulated through the point-spread function.
Technical Data, Wavefront
Measuring range sphere: –20 D to +20 D
Axis 0° – 180°
Measuring surface 2.0 mm to 7.0 mm (3 zones)
No. of measuring points up to 1500
Method Hartmann-Shack
Reference wavelength 1 555 nm (ISO 24157)
Technical Data, Corneal Topography
No. of rings 22 (18 complete rings)
No. of measuring points 3,425
Detected corneal surface
at 42.125 D Dia. 0.75 mm to 9.4 mm
Dioptres Measurement range 25 to 65 D
Accuracy ± 0.05 D (± 0.01 mm)