Analyzer CA-660, production date: 02/2016, power supply: 100-240VAC, power: 300VA. Visual analyzer in very good condition (5 mm paint chip on the upper surface of the housing - above the display). Inspection valid until 26.09.2023. The electronics start up correctly. English language. A message appears about the need to calibrate the LED. The Ready label appears on the screen. The time has been corrected. Counter status: 0/0/0 - after the first switch-on, the analyzer showed the Initialize the settings screen as during the original installation, OUS Setting was selected. The correct operation of: main fan, printer, stirrer, drawer lock, display, touch panel was checked. Checking selected sensors: stop, cover, probe crash, liquid sensor, sample rack. Carriage secured with cardboard filling the space. Handheld barcode reader (Honeywell, model: 1900GSR-1, SN: 15186B1508) - tested-functional. User manual not included.