Product Details
| Model | BFT II |
| Brand | SIEMENS |
| Category | Coagulation Analyzer |
| Posted | 26.01.2026 |
| Views | 268 |
Used - Good SIEMENS BFT II
Condition:: Used - Good
Availability:: In stock
Shipping from::
Poland
Safe deal
Bimedis secures your payment until you confirm the equipment is received.
Description
Siemens, Coagulology, BFT II, SN: B0281178
2 new magnets were glued into the reagent mixer. Upper housing after gluing and after painting. Lower case unglued in original green colour. 4 missing feet of housing completed. Analyser starts correctly, printer works, display works, keyboard works. Maintenance of the interior and exterior has been performed. Service tests and ‘Utylites’ check carried out - see below. Analyser operational for the activities performed, in good condition. ‘Wet’ tests were not carried out.
1 ‘Utilities’ settings: Printer-ON (parameter-protocol/OFF), Beeper-ON (OFF/Click), Reagent-stirrer-ON (OFF), Secret number-11111, Cuvette-test-OK.
2 Service tests: Optic: 104-103, 104-108 / Mixer: 375 OK / Stirrer-motor - OK / Heating - OK / Keyboard - OK / Display - OK / Calib. Temperature - measurement 37.5 (item 30, water, mercury, expected 37.4)
Set composition:
1. Fibrintimer II (BFT*II)
2. thermal printer paper roll
3. paper holder (set, 2+1 items) - alternative
4) Power cable 230VAC.
5. NONE: Reagent bottle adapter GW5 for 5 ml reagent bottle, should be 3 pcs.
6. NONE: Measuring port caps 2 pcs.
7. NONE: 100 cuvettes (Dispo-System), Teflon reagent mixing paddle, equipment cover, spare fuses, instruction manual.
2 new magnets were glued into the reagent mixer. Upper housing after gluing and after painting. Lower case unglued in original green colour. 4 missing feet of housing completed. Analyser starts correctly, printer works, display works, keyboard works. Maintenance of the interior and exterior has been performed. Service tests and ‘Utylites’ check carried out - see below. Analyser operational for the activities performed, in good condition. ‘Wet’ tests were not carried out.
1 ‘Utilities’ settings: Printer-ON (parameter-protocol/OFF), Beeper-ON (OFF/Click), Reagent-stirrer-ON (OFF), Secret number-11111, Cuvette-test-OK.
2 Service tests: Optic: 104-103, 104-108 / Mixer: 375 OK / Stirrer-motor - OK / Heating - OK / Keyboard - OK / Display - OK / Calib. Temperature - measurement 37.5 (item 30, water, mercury, expected 37.4)
Set composition:
1. Fibrintimer II (BFT*II)
2. thermal printer paper roll
3. paper holder (set, 2+1 items) - alternative
4) Power cable 230VAC.
5. NONE: Reagent bottle adapter GW5 for 5 ml reagent bottle, should be 3 pcs.
6. NONE: Measuring port caps 2 pcs.
7. NONE: 100 cuvettes (Dispo-System), Teflon reagent mixing paddle, equipment cover, spare fuses, instruction manual.