The electronics connections were checked and the main shaft responsible for the carriage movement was maintained. No resetting was observed after these activities. The analyzer (newer type) is in very good visual condition. Power supply 90- 264VAC, 145VA. Software: V02.02.03 (c) Aug 19 2005. Factory painted housing: top lighter, bottom slightly darker. The analyzer starts correctly, the printer is functional, the display is functional, the keyboard is functional. The external surfaces were maintained. Service tests and the Utilities inspection were carried out - see below. Wet tests were not carried out.
Kit composition:
1. Fibrintimer II (BFT II Analyzer)
2. Thermal printer paper roll
3. Paper holder (original set, 2 elements)
4. 230VAC power cable.
5. GW5 reagent bottle adapter for 5 ml reagent bottle (h23xfi22.4xfi30.3), 3 pcs.
6. Measuring socket caps 2 pcs.
7. NO: 100 cuvettes (Dispo-System), Teflon reagent mixing paddle, equipment cover, spare fuses, user manual.
1. Utilities settings: Printer-ON (parameter-protocol/OFF), Beeper-ON (OFF/Click), Reagent-stirrer-ON (OFF), Secret number-11111, Cuvette-test-133/579-132/635, cuvette detection-2xOK.
2. Service tests: Computer-OFF(Research-PC/Data-PC), Mathe-???, Optic: 122-112, 121-124 / Mixer: 375 OK / Stirrer-motor – OK / Heating – OK / RS232-???, Keyboard – OK / Display – OK / Calib. Temperature – measurement 37.5 (position 30, water, mercury, expected 37.4).