Product Details
| Model | Pentra C400 |
| Brand | HORIBA MEDICAL |
| Category | Biochemical Analyzers |
| Posted | 16.02.2026 |
| Views | 275 |
Product Specifications
- Analizator type
- Automatic
- Cycle time (reaction), s
- 20
- Sample Types
- Serum, Plasma, Urine, CSF, Whole Blood, Homogeneous Fluids
- Sample size, microlitre
- 0.7, 4, 5, 7 and 10
- Throughput, samples per hour
- 420
- Type of analyzer
- open
- Wavelength ranges, nm
- 340-755
Product Video
Used HORIBA MEDICAL Pentra C400
Condition:: Used
Shipping from::
USA
Safe deal
Bimedis secures your payment until you confirm the equipment is received.
Description
From routine to specific tests, perform them all on a single system
55 on-board assays
Innovative cassette format
Integrated work station and validation station
Touch screen
Real-time viewing of your workflow
Specifications:
Physical
System Size: 101 x 71 x 63cm (W x L x H)
System Weight: 120 kg
External Cooling Unit Size: 29 x 42 x 40cm (W x L x H)
External Cooling Unit Weight: 35 kg
Sound level: < 60 dBa
Operating temperature: from 15 to 32°C
Operating Humidity: from 20 to 85%
Total Power Requirements: 1000 VA
Water: De-ionized/Distilled Water – 10L Tank
Water specifications: Resistivity > 5 MegOhms, Conductivity < 0.2 µS/cm
Computer & Software
Operating system: Windows 10
Data processing:
12 inch Color LCD Touch Screen
Database Capacity: 100,000 results
Processor: 600 MHz
RC 232C: Patient results, QC to LIS with ASTM protocol
Quality Control Management: Levey Jennings, Automatic QC, Default controls and Westgard rules
Logs:
Reagents, Quality Control, Calibrations
Applications, Maintenance, Sequencing, Flags and Alarms
Export of archived data and Quality Control data (QCP): USB key
Interactive Help: HTML Format
Optical System
Measurement Principle:
Bi-chromatic light absorbance
Choice of 15 wavelengths from 340 to 700 nm
Wavelength selection with holographic grating technology
Sensor: Photo diode array
Optical Linearity: 2.95 A. at 340nm; 3.0 at other wavelengths
55 on-board assays
Innovative cassette format
Integrated work station and validation station
Touch screen
Real-time viewing of your workflow
Specifications:
Physical
System Size: 101 x 71 x 63cm (W x L x H)
System Weight: 120 kg
External Cooling Unit Size: 29 x 42 x 40cm (W x L x H)
External Cooling Unit Weight: 35 kg
Sound level: < 60 dBa
Operating temperature: from 15 to 32°C
Operating Humidity: from 20 to 85%
Total Power Requirements: 1000 VA
Water: De-ionized/Distilled Water – 10L Tank
Water specifications: Resistivity > 5 MegOhms, Conductivity < 0.2 µS/cm
Computer & Software
Operating system: Windows 10
Data processing:
12 inch Color LCD Touch Screen
Database Capacity: 100,000 results
Processor: 600 MHz
RC 232C: Patient results, QC to LIS with ASTM protocol
Quality Control Management: Levey Jennings, Automatic QC, Default controls and Westgard rules
Logs:
Reagents, Quality Control, Calibrations
Applications, Maintenance, Sequencing, Flags and Alarms
Export of archived data and Quality Control data (QCP): USB key
Interactive Help: HTML Format
Optical System
Measurement Principle:
Bi-chromatic light absorbance
Choice of 15 wavelengths from 340 to 700 nm
Wavelength selection with holographic grating technology
Sensor: Photo diode array
Optical Linearity: 2.95 A. at 340nm; 3.0 at other wavelengths